Agilent Technologies InfiniiVision 7000B Series Guide de l'utilisateur Page 141

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 428
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 140
Triggering 4
InfiniiVision Oscilloscopes User’s Guide 141
When triggering on a positive pulse, the trigger will occur on the high to low transition
of the pulse if the qualifying condition is true. When triggering on a negative pulse, the
trigger will occur on the low to high transition of the pulse if the qualifying condition is
true.
4 Press the qualifier softkey (< > ><) to select the time qualifier.
The Qualifier softkey can set the oscilloscope to trigger on a pulse width that is:
less than a time value (<).
For example, for a positive pulse, if you set t<10 ns:
greater than a time value (>).
For example, for a positive pulse, if you set t>10 ns:
within a range of time values (><).
For example, for a positive pulse, if you set t>10 ns and t<15 ns:
5 Select the qualifier time set softkey (< or >), then rotate the Entry knob to set the
pulse width qualifier time.
The qualifiers can be set as follows:
2 ns to 10 s for > or < qualifier (5 ns to 10 s for 350 MHz bandwidth models)
10 ns to 10 s for >< qualifier, with minimum difference of 5 ns between upper and
lower settings
< qualifier time set softkey
When the less than (<) qualifier is selected, the Entry knob sets the oscilloscope to
trigger on a pulse width less than the time value displayed on the softkey.
Trigger
10 ns
10 ns
Trigger
10 ns10 ns
Trigger
12 ns
15 ns
10 ns
Vue de la page 140
1 2 ... 136 137 138 139 140 141 142 143 144 145 146 ... 427 428

Commentaires sur ces manuels

Pas de commentaire