Agilent Technologies InfiniiVision 7000B Series Guide de l'utilisateur Page 211

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 428
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 210
Measurements and Math Functions 6
InfiniiVision Oscilloscopes User’s Guide 211
Overshoot and Preshoot Measurements
Preshoot
Preshoot is distortion that precedes a major edge transition expressed as a percentage of
Amplitude. The X cursors show which edge is being measured (edge closest to the trigger
reference point).
Digital channel time measurements
Automatic measurements Preshoot and Overshoot are not valid measurements for the math FFT
function or for digital channels on the mixed-signal oscilloscope.
Rising edge preshoot =
Base - D local Minimum
Amplitude
-----------------------------------------------------------
X 100
Falling edge preshoot =
local Maximum - D Top
Amplitude
----------------------------------------------------------
X 100
local Maximum
local Minimum
Preshoot
Preshoot
Top
Base
Vue de la page 210
1 2 ... 206 207 208 209 210 211 212 213 214 215 216 ... 427 428

Commentaires sur ces manuels

Pas de commentaire