Initial Setup 2
InfiniiVision Oscilloscopes User’s Guide 67
When the horizontal mode is set to Normal, the horizontal position control (delay control)
positions the normal sweep relative to the trigger. This delay is a fixed number. Changing
this delay value does not affect the sweep speed. When the horizontal mode is set to
Zoom, the delay controls the position of the zoom sweep window inside the normal
sweep display. This delay value is independent of sampling interval and sweep speed.
To change the sweep speed for the normal sweep window, press the Normal softkey,
then turn the sweep speed knob.
Time/div for
normal
sweep
Time/div for
Zoom window
Delay time
momentarily displays
when the Horizontal
Position knob is turned
Select Normal
sweep or Zoom
Zoom
window
Normal
sweep
window
These markers define the
beginning and end of the
Zoom window
Signal
anomaly
expanded
in zoom
window