Performance Tests Power Tests
108 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Test Equipment - Power Stability Tests
Figure 33 Test Setup for Power Stability Tests
Test Procedure, Low SSE output
This test procedure is applicable to the:
It measures the power stability of the module at the outputs indicated.
1 Set up the equipment as shown in Figure 23.
2 Connect the fiber to the low SSE output, Output 1.
Set <Optical Output> to <Low SSE (1)>.
3 Move to the TLS channel of the Agilent 8164A/B Lightwave
Measurement System and set the [Menu] parameters to:
Agilent 81600B #200 (Output 1, Low SSE)
Agilent 81600B #160 (Output 1, Low SSE)
Agilent 81600B #150 (Output 1, Low SSE)
Agilent 81600B #140 (Output 1, Low SSE)
Agilent 81600B #130 (Output 1, Low SSE)
Tunable Laser Channel [Menu] Parameters Values
<Wavelength Mode><λ>
<Source State><Off>
<Power Unit><dBm>
<Power Mode><Manual Att>
Modulation: <mod src><Off>