Test Re cords Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 195
Agilent 81600B #140 Performance Test Page 5 of 14
Model Agilent 81600B #140 Tunable Laser Report No. ________ Date_______
Mode Hop Free Tuning
1
Wavelength Deviation = Wavelength Measured - Wavelength Setting
Mode Hop Free Tuning Result (= Largest value of either the Maximum or Minimum Deviation)
Mode Hop Free Tuning Result: _______ nm
Te s t L i mi t : ±0.025 nm
Measurement Uncertainty: ±0.2 pm
Wavelength Setting Wavelength Measured Wavelength Deviation
1
1420.000 nm nm nm
1421.000 nm nm nm
1422.000 nm nm nm
1423.000 nm nm nm
1424.000 nm nm nm
1425.000 nm nm nm
1426.000 nm nm nm
1427.000 nm nm nm
1428.000 nm nm nm
1429.000 nm nm nm
1430.000 nm nm nm
1460.000 nm nm nm
1461.000 nm nm nm
1462.000 nm nm nm
1463.000 nm nm nm
1464.000 nm nm nm
1465.000 nm nm nm
1466.000 nm nm nm
1467.000 nm nm nm
1468.000 nm nm nm
1469.000 nm nm nm
1470.000 nm nm nm
Maximum Deviation: nm
Minimum Deviation: nm