Agilent Technologies Agilent 8164A Guide de l'utilisateur Page 127

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 272
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 126
Power Tests Performance Tests
Agilent 81600B Tunable Laser Source Family, Fourth Edition 127
10 Determine the limits of the transmission and attenuation ranges:
a Lower Transmission Band: λ
1
... λ
2
•OSA_λ
1
= OSA_λ
0
- 15 nm
•OSA_
λ
2
= OSA_λ
0
- 0.5 x Attenuation Band
= OSA_λ
0
- 1 nm
b Upper Transmission Band: λ
3
... λ
4
•OSA_λ
3
= OSA_λ
0
+ 0.5 x Attenuation Band
= OSA_λ
0
+ 1 nm
•OSA_
λ
4
= OSA_λ
0
+ 0.5 x Upper Transmission Band
= OSA_
λ
0
+ 15 nm
c Note the values of OSA_λ
1
, OSA_λ
2
, OSA_λ
3
, OSA_λ
4
in the Test
Record.
•OSA_λ
1
= ___________ nm
•OSA_
λ
2
= ___________ nm
•OSA_
λ
3
= ___________ nm
•OSA_
λ
4
= ___________ nm
11 Determine the SSE power values inside the transmission bands:
a Ensure the TLS is set to TLS_λ
0
and is not changed.
b Set the OSA marker to OSA_λ
1
c Check the OSA and note the SSE power value in [pW] in the table as
SSE_power
d Increase the OSA marker wavelength by 1 nm
e Repeat step c and step d until the wavelength is equal to OSA_λ
2
f Set the OSA to OSA_λ
3
g g Repeat step c and step d until the wavelength is equal to OSA_λ
4
.
h Add up all power values inside the transmissions bands to get the
value of power_trans.
Record the "partial noise power level" values in a table in [pW], where
1 pW = 10
-12
W.
Vue de la page 126
1 2 ... 122 123 124 125 126 127 128 129 130 131 132 ... 271 272

Commentaires sur ces manuels

Pas de commentaire