Agilent Technologies Agilent 8164A Guide de l'utilisateur Page 212

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 272
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 211
Performance Tests Test Records
212 Agilent 81600B Tunable Laser Source Family, Fourth Edition
Agilent 81600B #130 Performance Test Page 8 of 12
Model Agilent 81600B #130 Tunable Laser Report No. ________ Date_______
Power Linearity Output 1, Low SSE
Maximum Power Linearity at current setting ______ dB
Minimum Power Linearity at current setting ______ dB
Total Power Linearity = (Max Power Linearity - Min Power Linearity) ______ dBpp
Upper Test Limit 0.2 dBpp
Measurement Uncertainty Using 81624A/B #C01 ± 0.02 dB
Using 81623A/B #C01 ± 0.025 dB
Power Linearity Output 2, High Power
Upper Power Levels
Maximum Power Linearity at current setting ______ dB
Minimum Power Linearity at current setting ______ dB
Total Power Linearity = (Max Power Linearity - Min Power Linearity) ______ dBpp
Upper Test Limit 0.6 dBpp
Measurement Uncertainty Using 81624A/B #C01 ± 0.02 dB
Using 81623A/B #C01 ± 0.045 dB
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity at
current setting
Start = REF - 6.0 dBm 0.00 dB + 0.00 dB = 0.00 dB
- 7.0 dBm dB + 1.00 dB = dB
- 8.0 dBm dB + 2.00 dB = dB
- 9.0 dBm dB + 3.00 dB = dB
- 10.0 dBm dB + 4.00 dB = dB
- 11.0 dBm dB + 5.00 dB = dB
- 12.0 dBm dB + 6.00 dB = dB
- 13.0 dBm dB + 7.00 dB = dB
Power Setting
from start
Measured Relative
Power from start
Power reduction
from start
Power Linearity at
current setting
Start = REF + 4.0 dBm 0.00 dB + 0.00 dB = dB
+ 3.0dBm dB + 1.00dB = dB
+ 2.0dBm dB + 2.00dB = dB
+ 1.0dBm dB + 3.00dB = dB
0.0 dBm dB + 4.00 dB = dB
- 1.0 dBm dB + 5.00 dB = dB
- 2.0 dBm dB + 6.00 dB = dB
- 3.0 dBm dB + 7.00 dB = dB
Vue de la page 211
1 2 ... 207 208 209 210 211 212 213 214 215 216 217 ... 271 272

Commentaires sur ces manuels

Pas de commentaire