Agilent Technologies Option H48 Multiport Test Set Z5623A Spécifications Page 144

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 191
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 143
6
Page 11
安捷倫科技高頻元件量測研討會
Feb.23, 2006
New Calibration Paradigm
Cal Registers
Each channel has its own cal register
All cal data is saved to the channel cal register
Data in cal register is “temporary” –
data is saved on hard drive, BUT will be
overwritten by the next calibration in that channel
User Cal Sets
Old “Cal Sets” are now called “User Cal Sets”
User must manually save data to a User Cal Set
User cal sets are stored on the hard drive as before
You no longer have a choice
to save an instrument state!
Page 12
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Cal Registers Should Be Familiar
Cal register model is similar to that used in 8720s and 8510s
Cal registers prevent accumulation of unneeded cal sets
Vue de la page 143
1 2 ... 139 140 141 142 143 144 145 146 147 148 149 ... 190 191

Commentaires sur ces manuels

Pas de commentaire