Unknown Thru Calibration Requirements
Systematic errors of all test ports (directivity, source match,
reflection tracking) can be completely characterized
“Unknown thru” calibration standard:
Must be reciprocal (Sij = Sji)
Phase known to within a quarter wavelength
VNA signal-path switch errors can be quantified
Requires dual reflectometers on all ports
(e.g., a 2-port 4-receiver VNA) OR
Requires delta-match correction
Page 48
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Two-Port Unknown Thru Calibration Sequence
Measure open, short, load on port 1
Measure open, short, load on port 2
Measure insertable adapter (unknown thru)
between ports 1 and 2
Confirm estimated electrical
delay of unknown thru
1-port calibrations
Unknown thru
Unknown Thru calibration is performed
just like a “flush thru” 2-port calibration!