Vue de la page 147
10
Page 19
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Agenda
Overview of PNA FW Rev. 6.0
Cal Registers
TRL Calibration with 4-port 20 GHz PNA-L
Unknown Thru Calibration
Offset Load Cal
Fixture And Probe Techniques
Auto port extensions
Embedding/de-embedding
Frequency Converter Applications – SMC
Measuring fixtures/probes
Q&A
Page 20
安捷倫科技高頻元件量測研討會
Feb.23, 2006
TRL Calibration with 4-port 20 GHz PNA-L’s
Yes, it can be done! (requires firmware A.05.25 and later)
Is “true” TRL calibration, not an approximation like TRL*
Two-step (two-tier) calibration process:
Step 1: “Delta Match” calibration
Step 2: Normal TRL calibration
What is a “Delta Match” cal?
characterizes the difference (delta) between the source match
and load match of each port of the VNA
allows a VNA having a single
reference receiver to
perform TRL-style calibrations (including Unknown-Thru)
Why 2 steps?