13
Page 25
安捷倫科技高頻元件量測研討會
Feb.23, 2006
Delta Match Cal
Delta match cal characterizes internal match differences,
so it can be performed
with mixed connector types
with or without test port cables, in any combination
with adapters that can remain or be removed for subsequent TRL cals
Page 26
安捷倫科技高頻元件量測研討會
Feb.23, 2006
When is Delta Match Used?
Definitions
Flush thru: insertable, zero-length
Known thru: characterized (all 4 S-parameters)
Unknown thru: reciprocal (S21=S12); usually low-loss and well matched
YesTRL - Adapter Removal
NoECal Internal Thru
YesECal Unknown Thru
YesTRL - Known Thru
YesSOLT - Unknown Thru
NoSOLT - Adapter Removal
NoSOLT - Known Thru
(Includes Flush-Thru, Known-Thru and Minimized Thru)
Uses Delta Match?Cal Method