Agilent Technologies 8000 Series Manuel de l'opérateur Page 27

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 126
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 26
Setting Up the 8000 3
8000 Operator’s Manual 27
3 Place the 806 Syringe Pump on a clean, dry, level section of the
bench top between the 8000 and the dissolution apparatus.
4 Refer to the following example diagram to correctly make cable
connections for the 8000, syringe pump, and dissolution apparatus:
Figure 6 Cable Connections for 8000, Syringe Pump, and VK 7000 /
7010
5 Connect the single end of the Y cable (#1, P/N 5075-0446) to the
RS232 port on the syringe pump.
6 Connect an RS232 cable (#2, P/N 5075-0252) between the RS232
COMPUTER port on the 8000 and one extension of the Y cable
attached to the syringe pump.
7 If the system configuration includes the 705-DS, connect a
9-to-9-pin cable (P/N 5075-0448) from the RS232 port on the
dissolution apparatus rear panel to the remaining extension of the Y
cable attached to the syringe pump.
Otherwise, connect a second RS232 cable (#2, P/N 5075-0252)
between the remaining extension of the Y cable attached to the
syringe pump and the RS232 port on the dissolution apparatus.
Vue de la page 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 125 126

Commentaires sur ces manuels

Pas de commentaire