64 8000 Operator’s Manual
4 Operating the 8000
6 Purge Volume (syringe
pump)
If your system configuration includes the syringe pump, use this option to set a purge
volume that ensures all stranded medium is properly expelled.
Select option 6, Purge Volume. Enter a value for the purge volume that ensures at
least two strokes of the syringe plunger and press ENTER. The Program Variables
screen displays.
Note: The first stroke moves all medium from the sampling lines to the return lines.
The second stroke draws air into the sampling lines and purges the remaining
medium from the return lines. As with priming, the longer the tubing lengths, the
larger the purge volume required.
7 R-M (peristaltic pump) If your system configuration includes the peristaltic pump, select option 7, R-M. The
Replacement Media Option screen displays. Select option 1, Enable. Enter the
desired percentage of replacement medium and press ENTER. Optionally, select
option 2, Disable. The Program Variables screen displays.
Note: If the RM option is not installed, select option 2, Disable.
7 Filter Chg (syringe pump
and filter changer)
If your system configuration includes the syringe pump and filter changer, select
option 7, Filter Chg. Enter the desired number of samples each filter should process
before being discharged and press ENTER. The recommended value is 1. The
Program Variables screen displays.
Note: If your system configuration includes the syringe pump and filter changer,
ensure the syringe pump option is enabled. From the Ready screen, press MENU.
Select option 3, Set Syringe Pump. Select option 1, Enable. Also press MENU > 0 >
P and select option 1, Enable.
The replacement media option is not available with this system configuration. Ensure
the RM option for the syringe pump is disabled using the hidden key function MENU
> 0 > R.
7 R-M (syringe
pump)
If your system configuration includes the syringe pump but not the filter changer,
select option 7, R-M. The Replacement Media Option screen displays. Select option
1, Enable, or option 2, Disable. If the RM option is not installed, select option 2,
Disable. The Program Variables screen displays.
Note: If your system configuration includes the syringe pump, ensure the syringe
pump option is enabled. From the Ready screen, press MENU. Select option 3, Set
Syringe Pump. Select option
1, Enable. Also press MENU > 0 > P and select option
1, Enable.
If your system configuration includes the syringe pump but not the filter changer, use
the hidden key function MENU > 0 > R to enable or disable the RM option, as
appropriate. If the RM option is not installed, select option 2, Disable.
Table 1 0 Program Variables Options (Apparatus 1 / 2)
Option Function