Setting Up the 8000 3
8000 Operator’s Manual 41
10 Load the new tubes of filters into the carousel of the filter changer.
See “Loading Filters (Filter Changer only)” on page 40.
11 Select option 3, Change Filters.
12 Enter 8 as the number of filters to change and press ENTER. The
filters expel from the filter changer.
13 When the filter changer stops expelling filters, select option 3,
Change Filters, again.
14 Enter 1 as the number of filters to change and press ENTER. The
remaining filter of the previous type expels from the filter changer
and is replaced with the new filter.
15 Press ESC four times to return to the Ready screen.
Hidden Key Functions
Table 1 Hidden Key Functions
Key Sequence Function
MENU > 0 > 0 Use this key sequence to select the system configuration. Select the option that
corresponds to the number of valves on your 8000.
MENU > 0 > 8 Use this key sequence to select a fixed 1200 baud or a programmable baud. Select
option 2, Programmable Baud.
Note: If you are using the 8000 with other Agilent equipment, the baud rate must be
set to 9600 baud (see Printer Operation and Communicationson page 103).
MENU > 0 > A Use this key sequence to set the drop volume before sampling. Enter the desired drop
volume and press ENTER. The maximum acceptable value is 1.
Note: The drop volume is an amount of sample that is dispensed through the 8000
needles prior to dispensing samples into the collection tubes to ensure the needles are
purged completely.
MENU > 0 > B Use this key sequence to enable or disable the moving alert. The asterisk (*) displays
next to the current selection. If enabled, the alarm sounds when the dispensing arm is
in motion.
MENU > 0 > C Use this key sequence to enable or disable the delayed start alarm. The asterisk (*)
displays next to the current selection. If the delayed start alarm is enabled, the alarm
sounds prior to the start of the program. See “DISSO Options Screen” on page 69.