Setting Up the 8000 3
8000 Operator’s Manual 39
External Unit
Figure 13 External Media Replacement Module
1 Remove the cover housing the RM chamber from the side of the
8000.
2 Detach the RM chamber.
3 Insert the end of the tubing attached to the IN fitting on the pump
into the plastic RM chamber.
4 Slide a tubing clamp on a length of tubing.
5 Connect one end of the tubing containing the clamp to the OUT
fitting on the pump. Connect the other end to the IN fitting of the RM
chamber.
6 Connect a length of plastic tubing to the OUT fitting of the RM
chamber.
7 Place the free end of this piece in the plastic RM chamber.
8 Reattach the RM chamber, tighten all screws, and replace all covers.