Agilent Technologies DP235 Guide de l'utilisateur Page 116

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 602
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 115
Contents 116 Index
The four peak find setpoints that can be changed are:
Min Peak Height The Min Peak Height setpoint determines whether a peak is kept.
For each peak, the difference between the start point value and the
center point value (local baseline) must be greater than the Min
Peak Height value.
Min Peak Width The Min Peak Width setpoint determines the minimum amount of
time that must elapse before a peak is recognized.
Slope Threshold The Slope Threshold setpoint determined the difference in the
slope that must occur in order for a peak to begin. The inverse of
this value is used to determine the peak end.
Baseline Plateau The Baseline Plateau setpoint is a parameter that assists in finding
peaks. The signal is recognized to be at baseline whenever the
slope of the data is less than the Slope Threshold setpoint (either
positive or negative) for longer than the time set for the Baseline
Plateau. This setting rejects brief, low slope areas such as between
non-baseline-resolved peaks.