How to Prepare the Bioanalyzer for Flow Cytometric Assays
Remove the electrode cartridge:
1 Turn off line power to the Agilent 2100 bioanalyzer. The line switch is located at the rear.
2 Open the lid.
CAUTION
Do not touch the electrodes while the cartridge is in the Agilent 2100 bioanalyzer. The
electrodes and the high voltage power supplies can be damaged.
3 Pull down the metal locking lever in the open position as shown in the figure below.
The cartridge is pushed out.
4 Gently pull the cartridge out of the lid.
Metal lever
in open
position