Performing Qualifications
To ensure a validated Agilent 2100 bioanalyzer system, qualification steps have to be
performed at installation and operation level.
2100 expert allows for detailed installation qualification (IQ) and operational qualification
(OQ) on both the bioanalyzer hardware and software. Each qualification comprises a
series of tests and measurements that you can run and document in the Validation
context of the 2100 expert software.
Installation Qualification
Installation qualification includes tests to verify that the bioanalyzer software and
hardware are installed properly and that all electrical and pressure connections are
correct.
Installation qualification must be performed once after installation.