Lab Exercise: Overview and Diagnostics
Diagnosis View
57
maintenance CD-ROM and then learn to perform user maintenance procedures on
the 1200 modules.
1) Insert the HPLC Maintenance and Repair CD into the CD-ROM drive. (The
CD may already be there.)
2) When the contents of the CD-ROM are displayed, double-click the file
start.htm. Alternatively, you can get started by clicking the Windows Start
button, selecting Run…, then Browse to find the CD-ROM drive. View
Files of type: All Files, select start.htm, then Open, and OK. The program
has a routine to check for necessary software components on your PC, which
should already be present. Click the Start Application button when it
appears.
3) The CD-ROM will now load. The Table of Contents appears in the left-
hand part of the window. Find the chapter on the Quaternary Pump and
click to expand its contents. Next find Repairing the Pump, and expand its
contents to see the Simple Repair Procedures. Simple repairs are those
that can be done from the front of the module without removing the module
from the instrument stack.
4) Under the Simple Repair Procedures, you will see a list of procedures.
The ones that have a video recorder icon next to them have a video clip as
part of the repair procedure. For practice, select the procedure for
Exchanging the Pump Seals and Seal Wear-in Procedure.
5) The instructions for this procedure will appear on the right side of the
program window. The information provided includes: When required,
Tools required, Parts required, Preparation and Completion Steps, and
the step by step directions. You can scroll down and read all the directions
or you can run the video clip by clicking the start button at the lower left side
of the video area. Once started, the start button becomes a pause button.
Buttons to stop the video clip and to adjust the volume are also present.
6) Note: You must have speakers or the headphones installed in order to use
the audio portion of the presentation.
7) Now try to find the procedure for changing the needle assembly in the
autosampler and play the video for this maintenance procedure.
NOTE: Some of the maintenance procedures will require you to use the
ChemStation software to move a part of the instrument. For instance, to change
the needle, the needle position must be moved up and down. In order to perform
this function, you must go to the Diagnosis view, select the Maintenance menu,
then ALS, and Maintenance Positions. When you are ready to begin the
maintenance procedure, you would select Start in the Change Needle box.
Changing the piston, changing the gripper and parking the arm for transport are
also performed from this window.