Lab Exercise: Sequence Review and Reprocessing
Reprocess with Sequence Folder Method
112
You will update the Sequence Folder Method with new retention time windows
and drop the 4
th
level calibration data point, then reprocess the entire sequence.
Note that only the method (.M) and sequence file (.S) in the sequence folder will
change. The master method stored in Chem32\1\Methods and the sequence
template stored in Chem32\1\Sequence will not change.
1) Go to Data Analysis.
2) In the Data Explorer, find the sequence folder: 03-20-06 and expand. Find
the sequence folder: Chem32\1\Data\03-20-06\SequenceClass 2006-03-20
13-48-40. Double-click to load into the Navigation Table.
3) Load the container method. The DA Method for the first file will load
automatically as stated in the Preferences. In this part of the exercise,
however, you are changing the folder method so the changes apply to all
data files. The path for the sequence folder method is: Chem32\1\Data\03-
20-06\SequenceClass 2006-03-20 13-48-40\CLASSQUANT.M. Load the
sequence folder method now.