Lab Exercise: Quantification
Preparations
86
5) To turn the solvent delivery system on, you may use the menu items:
Instrument, More Pump, Control; or access the tool shown in the diagram
below. Do this now.
6) Wait until a steady stream of solvent comes out of the purge valve waste tube.
7) Repeat steps 3 through 6 for the other channels of the pump. Complete the
priming process by setting the composition to %B=65 and pump at 5 mL/min.
8) Change the flow rate to 1.5 mL/min and then close the purge valve. Allow the
column to equilibrate.