Creating a rise time testing mask
Mask template testing can be used to test the rise time of a signal, including
specifying an upper limit for rise time. For example, you can specify that the
rise time must be 15 ns or faster to pass the test.
Use the voltage and time readouts of the mask editor to ensure the correct
settings. In the following figure, T1 and T2 are the critical points for
determining the maximum rise time limit (rise time limit = T2 – T1).
Example of a definition of a rise time testing mask
Figure 2 16
Operating the Measurement/Storage Module