Module 6
Low Current Measurement
6-28
Maximum resolution
16 PLC
Gummel Plot
Range and Integration Time
Medium integration smoothes the plot at the ultra low levels. Make sure the range is set to AUTO or
LIMITED to lowest current range:
HRSMU: 10 pA
MPSMU: 1 nA
HPSMU: 1 nA