Module 5
Basic Measurement
5-8
Force Lines to Fixture 1,2,3,4 Ports
16442A/B Fixture
B1500A Rear View
SMU4
SMU3
SMU2
SMU1
CMU
GNDU
Force side
Sense side
SMU Triax Connection
For the non-Kelvin connections, connect only the Force lines, leaving the Sense lines open.
Connect corresponding numbers. On the 16442A/B fixture use the numbers labeled 1 - 6, not 1 - 3.
Your B1500A may not match the SMU configuration shown in this figure. Note that SMU1 is the
module top of the GNDU (ground unit). The SMU number become large from bottom to top as
shown.
This is the SMU cable setup that will be used in the remainder of the class exercises.