136 Programmer’s Guide
5 Command Reference
CAT Select A Cable
[:SENSe]:CAT:CABLe:LOAD <string>
Use this command to select a cable from built-in cable list.
Example: :CAT:CABLe:LOAD "5088"
Key access: Meas > More 1 of 2 > Cable Specification > Select Cable
CAT Cable Attenuation
[:SENSe]:CAT:CABLe:ATTenuation <value>
[:SENSe]:CAT:CABLe:ATTenuation?
This command defines the attenuation value for a custom cable.
Range: 0 to 5 dB/m, or 0 to 1.524 dB/ft
Example: :CAT:CABLe:ATTenuation 1.2
Key access: Meas > More 1 of 2 > Cable Specification > Cable Type > Custom, then
Cable Atten
CAT Cable Velocity Factor
[:SENSe]:CAT:CABLe:VELFactor <value>
[:SENSe]:CAT:CABLe:VELFactor?
This command defines the velocity factor for a custom cable.
Range: 1 to 100
Example: :CAT:CABLe:VELFactor 89.2
Key access: Meas > More 1 of 2 > Cable Specification > Cable Type > Custom, then
Vel Factor