Vue de la page 152
Command Reference 5
Programmer’s Guide 145
CAT Limit Type
:CALCulate:LLINe[1]|2:TYPE UPPer|LOWer
:CALCulate:LLINe[1]|2:TYPE?
This command toggles the limit type between upper limit and lower
limit.
Example: :CALCulate:LLINe:TYPE UPPer
Key access: Shift > Limit > Limit1 > Type
CAT Limit Line State
:CALCulate:LLINe[1]|2:STATe OFF|ON|0|1
:CALCulate:LLINe[1]|2:STATe?
This command turns on/off the limit line in CAT test.
Example: :CALCulate:LLINe:STATe ON
Key access: Shift > Limit > Limit1 > Limit Line On/Off
CAT Limit Line Value
:CALCulate:LLINe[1]|2:Y <value>
:CALCulate:LLINe[1]|2:Y?
This command defines the limit line value in cable and antenna
measurement mode.
Example: :CALCulate:LLINe:Y 15
Key access: Shift > Limit > Limit1 > Limit Line On/Off
CAT Limit State
:CALCulate:LLINe[1]|2:DISPlay OFF|ON|0|1
:CALCulate:LLINe[1]|2:DISPlay?
This command turns on/off the limit in cable and antenna measurement
mode. This limit is different from the limit line describe above, it is
combined by several limit line with different amplitudes and different
frequency points.
Example: :CALCulate:LLINe1:DISPlay 1
Key access: Shift > Limit > Limit1 > Limit On/Off