208 Programmer’s Guide
5 Command Reference
Set Bottom Value of Display Range
:DISPlay[:WINDow]:DISPrange:BOTTom <value>
:DISPlay[:WINDow]:DISPrange:BOTTom?
This command defines the bottom value of display range
Range: –200 to +199.99 dBm
Example: :DISPlay:WINDow:DISPrange:BOTTom 20dBm
Key access: Meas > Meas Disp > Disp Range > Bottom
Display Mode
:DISPlay[:WINDow]:MODE METer|CHARt|TRACe
:DISPlay[:WINDow]:MODE?
This command to toggle display mode between meter, chart and trace.
Example: DISPlay:WINDow:MODE CHARt
Key access: Meas > Meas Disp > Disp Mode
Set Chart View to Home
:DISPlay[:WINDow]:CHARt:HOME
This command set the chart view page to home.
Example: :DISPlay:CHARt:HOME
Key access: Meas > Meas Disp > Chart View > Home
Set Chart View to End
:DISPlay[:WINDow]:CHARt:END
This command set the chart view page to end.
Example: :DISPlay:CHARt:END
Key access: Meas > Meas Disp > Chart View > End