Command Reference 5
Programmer’s Guide 149
Bottom N Number
:CHSCanner:SWEep:BOTTom <value>
:CHSCanner:SWEep:BOTTom?
This command specify the scanner channel number of Bottom N mode.
Need to switch to Bottom N sweep mode before using this command.
Range: 1 to 20
Example: :CHSCanner:SWEep:BOTTom 10
Key access: Meas > Channel Scanner > Scan Mode > Bottom N
Max Hold State
:CHSCanner:SWEep:MAXHold[:STATe] OFF|ON|0|1
:CHSCanner:SWEep:MAXHold[:STATe]?
Use this command to turn on/off the trace max hold state.
Example: :CHSCanner:SWEep:MAXHold ON
Key access: Trace > Max Hold
Max Hold Dwell Time
:CHSCanner:SWEep:MAXHold:DWELltime <value>
:CHSCanner:SWEep:MAXHold:DWELltime?
Use this command to define the max hold dwell time of channel scanner
measurement mode.
Range: 100 ms to 3600 s
Example: :CHSCanner:SWEep:MAXHold:DWELltime 10s
Key access: Trace > Max Hold
Mini Hold State
:CHSCanner:SWEep:MINHold[:STATe] OFF|ON|0|1
:CHSCanner:SWEep:MINHold [:STATe]?
Use this command to turn on/off min hold state.
Example: :CHSCanner:SWEep:MINHold ON
Key access: Trace > Min Hold