1100 LC Modules Control Manual - 158 - 03-914947-30:Rev.6
Perform the Pressure Test to demonstrate pressure–tightness up
to 400 bars. The test consists in monitoring pressure profile while
the pump is following defined stages.
After its initialization, the pump starts pumping isopropyl alcohol
with a flow rate of 510µL/min and a stroke of 100 µL. When the
system pressure reaches the 390s bar value, the pump switches
off.
Once the pressure test is finished, results are displayed on
screen and also in the maintenance audit trail: