1100 LC Modules Control Manual - 21 - 03-914947-30:Rev.6
Communications Parameter Group
Master module HPIB Address:
Select this option if the HPLC chain is controlled via a GPIB
communication and enter the GPIB address of the Master
module.
NOTE: To check the GPIB address value of the Master module, refer to the Appendix at
the end of this manual.
HPIB Hardware:
If the HPLC chain is controlled via a GPIB communication, select
the name of the bus previously configured in the HPIB Hardware
scrolling list.
NOTE: For more details about the Bus configuration, refer to the Configuring the
Communication Server section in this manual.
Master module IP Address:
Select this option if the chromatograph is controlled via Ethernet
communication and type the IP address of the instrument.
Miscellaneous Parameter Group
Check or uncheck the Galaxie run started by this module (No
controlled autosampler) check box. This box should be checked
if a non-controlled autosampler present in the system is cabled
to this module to give a start signal to the 1100 module. The
Galaxie run will start as soon as this module is started.
Hardware Parameter Group
Check the Switch valve option installed check box if a switch
valve is present in the module.