1100 LC Modules Control Manual - 169 - 03-914947-30:Rev.6
For analyzing the test results, refer to the detector operating
manual.
Dark Current Test
Perform the Dark Current test to determine the leakage current
from each optical diode. This leakage current may cause non-
linearity or excessive noise at specific wavelengths.
When starting the test, the slit 1 nm assembly is moved to the
dark position. At the end of the dark Current test, the resulting
leakage current is measured and graphically displayed for each
diode. Results are also listed in the maintenance audit trail: