1100 LC Modules Control Manual - 73 - 03-914947-30:Rev.6
In the Range scrolling list, enter the output voltage range value.
The available values are: 0.1 Volt Full Scale and 1Volt Full
Scale.
In the Source scrolling list, enter the collected signal type. The
available values are: Absorbance Units AU, Sample AU without
Reference, Reference AU without Sample, Test Chromatogram,
Unfiltered and not logarithm Sample, Unfiltered and not logarithm
Reference, Analog Output Test Function, Sample Diode Current,
Reference Diode Current, Lamp Anode Voltage, Lamp Anode
Current, Lamp Heater Voltage, Lamp Heater Current, Grating
Motor Step, Filter Motor Step, Board Temperature and PTC
current.
In the Polarity scrolling list, enter the detector polarity. The
available values are: Positive and Negative.
Spectra Control