Agilent Technologies 1100 Series Manuel de service Page 102

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 318
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 101
102 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Exchanging a Lamp
When required If noise or drift exceeds application limits or lamp does not ignite
Tools required Screwdriver POZI 1 PT3
Parts required Longlife Deuterium lamp 2140-0813 (without black cover)
Longlife Deuterium lamp 5181-1530 (with black cover)
Deuterium lamp 2140-0590 (without black cover)
Tungsten lamp G1103-60001
Preparations for this procedure:
Turn the lamp(s) off.
CAUTION:
If the detector has been in use, the lamp may be hot.
If so, wait five minutes until lamp cools off.
1
Press the release buttons and remove the front cover to
have access to the lamp area.
Vue de la page 101