152 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
5 To re-install the sensor holder (here shown with the
sensor), bring the holder into its position and slide it
carefully across the fan cable.
6 Clip the sensor carefully into its holder and route the
cables as shown.
Assure, that the fan rotates smoothly without any
scraping noise. Otherwise correct the position of the
sensor holder.
7 Position the heater assembly as shown and clip the heater
onto the fan. The cables should be oriented towards the
left corner.
The fan must be installed in the correct orientation
(see arrow on the fan) to ensure optimum cooling
and operation of the detector. The direction of air
flow is from the rear towards the front (see figure on
next page).