Agilent Technologies 1100 Series Manuel de service Page 55

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 318
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 54
1100 Series DAD and MWD Reference Manual 55
Troubleshooting and Test Functions 3
Shutdown
An external instrument has generated a shut-down signal (through CAN or
REMOTE lines).
The detector continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
Leak detected in an external instrument with a remote connection to the
system.
Shut-down in an external instrument with a remote connection to the
system.
The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
Fix the leak in the external instrument before restarting the pump.
Check external instruments for a shut-down condition.
Check the degasser for an error condition. Refer to the Reference Manual
for the Aligent 1100 Series vacuum degasser.
Vue de la page 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 317 318

Commentaires sur ces manuels

Pas de commentaire