Agilent Technologies 1100 Series Manuel de service Page 153

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 318
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 152
1100 Series DAD and MWD Reference Manual 153
Repairing the Detector 4
The figure below shows the required air flow
direction.
8 Insert the fan assembly into its location.
9 Reconnect the connectors to the main board (fan J16),
temperature sensor (J18) and heater (J17), see figure
below for location.
Next Steps:
Reinstall the top foam section, top cover and front cover,
see “Replacing the Foam and the Top Cover” on
page 175.
Turn on the detector and prove, that the fan rotates
smoothly without any scraping noise. Otherwise repeat
step 5 and step 6 on page 152.
Replace the detector into the stack.
Reconnect cables, capillaries and turn on detector
AIRFLOW
G1315/65A
G1315/65B
Vue de la page 152
1 2 ... 148 149 150 151 152 153 154 155 156 157 158 ... 317 318

Commentaires sur ces manuels

Pas de commentaire