150 Chapter 3
Screen Reference
CONFIGURATION
BER BS Level
b
Specifies the RF output level of the Test Set for the Sensitivity test.
The allowable range is from 110.0 to 50.0 dBm in 0.5 dB steps.
BER Frames
b
Specifies the number of test frames. The allowable range is from 1
to 13000. The number of test bits, derived from the equation “Test
Frames multiplied by 78 bits”, is shown in the parentheses.
Loopback Delay
Selects a delay time from Short, Mid, or Long.
a. In the AUTOMATIC TEST mode, the High value is used for the Burst
Timing, Power Ramp, Phase Error, and Frequency Error tests.
b. The variable increment softkey is activated. Refer to "Variable Incre-
ment Softkey" on page 60.