24 Chapter 1
Preparing for Use
Preparation for Each Test
Entering Attenuations
Determine and enter the appropriate attenuation values caused by the antenna
coupler, RF cable, or shield box that you use to connect the mobile phone under
test to the Test Set.
Step 1. By pressing the CONFIGURATION softkey, obtain the following
CONFIGURATION screen:
Step 2. Set the Attenuation field to On by rotating and pressing the CURSOR CONTROL
knob.
Step 3. Enter the appropriate loss values, depending on the radio systems, in the RF In and
RF Out fields with the CURSOR CONTROL knob. These values are also used in
other function modes.
Step 4. Press the Return softkey to reveal the initial screen with function modes.
Connecting a Printer (optional)
For obtaining screen hardcopies, connect a printer to the Test Set as follows using
an appropriate interface cable between the PRINTER connector on the rear panel
of the Test Set and the input connector of a printer:
Figure 1-4 Connecting a Printer