Agilent Technologies DP235 Guide de l'utilisateur Page 151

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 628
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 150
151
Using the Oscilloscope
Calibrating the oscilloscope
4 Calibrate the Logic Trigger of the Self Cal.
a Obtain a BNC 50-W, 48-inch cable.
b Select Start. The instrument will prompt you to connect the
cable to the appropriate location on the rear panel of the
instrument.
c Select the Procedure field, then select Logic Trigger from
the pop-up menu.
d Select the Start field and follow the instructions on the
display.
e After completion of the Logic Trigger calibration, remove the
cable from the instrument.
Protect the operational accuracy calibration
factors
Facing the front panel, reach around to the lower right corner on the back
of the logic analyzer. Flip the PROTECT/UNPROTECT switch to
PROTECT (flip the switch down).
Vue de la page 150
1 2 ... 146 147 148 149 150 151 152 153 154 155 156 ... 627 628

Commentaires sur ces manuels

Pas de commentaire