Using the Oscilloscope
Oscilloscope Common Menus
Oscilloscope Common Menus
The following options apply to all of the oscilloscope menus.
Run/Stop options
There are three ways you can manually run and stop the oscilloscope:
the Autoscale menu, the Run and Stop keys, and the Run/Stop field.
Single and Repetitive modes
Single mode acquisition fills acquisition memory once with 8000
samples of the input waveform, automatically stops running, then
displays the contents of acquisition memory. Each 8000-sample
waveform record is acquired in a single acquisition. Each channel has a
memory capacity of 8000 samples.
Repetitive mode acquisition fills acquisition memory with 8000 samples
of the input waveform on continuing acquisitions, with each new
acquisition overwriting the previous. The display is updated each time
a new acquisition is made. Repetitive mode continues acquiring data in
this manner until you select the Stop field. As in single mode, each
8000-sample waveform record is acquired in a single acquisition.
Autoscale run
Select the autoscale field on the screen, then choose Continue from the
pop-up menu. When autoscaling is complete, the oscilloscope
automatically starts running.