Agilent Technologies DP235 Guide de l'utilisateur Page 297

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 628
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 296
297
Logic Analyzer Reference
The Analyzer Format Menu
The Analyzer Format Menu
Pod threshold field
The pod threshold field is used to set a voltage level that the data must
reach before the analyzer recognizes and displays it as a change in logic
levels. Threshold levels apply to single pods, and cover both data and
clock channels.
TTL. When TTL is selected, the threshold level is +1.5 volts.
ECL. When ECL is selected, the threshold level is -1.3 volts.
User. When User is selected as the threshold level, the data signals
must reach a user selectable value. The range of this value is between -
6.0 volts to +6.0 volts in 50-mV increments.
Vue de la page 296
1 2 ... 292 293 294 295 296 297 298 299 300 301 302 ... 627 628

Commentaires sur ces manuels

Pas de commentaire