Agilent Technologies DP235 Guide de l'utilisateur Page 202

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 628
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 201
202
Using the Pattern Generator
Building Test Vectors and Functions
To build an initialization sequence
Use the INIT SEQUENCE to place the system under test into a known
initialization state. Default start and end program vectors are marked
INIT SEQUENCE START and INIT SEQUENCE END. During a
repetitive run, the initialization sequence is only executed the first time
the program is run. The main sequence then loops repetitively.
1 From the Sequence menu, use the knob to highlight INIT
SEQUENCE START.
2 Select the Insert field once for each new line of vectors you want
inserted into the initialization sequence.
These new lines of vectors provide fields to place data into for each
label.
3 Select the data field, then type in a data value.
If you are adding data in many fields, use the Autoroll feature. Refer to
"To use Autoroll" on page 217.
Vue de la page 201
1 2 ... 197 198 199 200 201 202 203 204 205 206 207 ... 627 628

Commentaires sur ces manuels

Pas de commentaire