4- 44 Agilent B1500A User’s Guide, Edition 13
Specifications
EasyEXPERT Software
Measurement modes
Supports the following measurement modes.
• IV measurement
Spot
Staircase sweep
Pulsed spot
Pulsed sweep
Staircase sweep with pulsed bias
Sampling
Multi-channel sweep
Multi-channel pulsed sweep
List sweep
Linear search
(*1)
Binary search
(*1)
• C measurement
Spot C
CV (DC bias) sweep
Pulsed spot C
(*1)
Pulsed sweep CV
(*1)
C-t sampling
(*1)
C-f sweep
(*1)
CV (AC level) sweep
(*1)
Quasi-static CV (QSCV)
(*1)
(*1)
Direct Control classic test is used.
Sweep measurement
Number of steps: 1 to 10001 (SMU), 1 to 1001 (CMU)
Sweep mode: linear or logarithmic (log)
Sweep direction: single or double
Hold time: 0 to 655.35 s, 10 ms resolution
Delay time: 0 to 65.535 s or 655.35 s for CV (AC level) sweep and C-f sweep, 100
s resolution