Agilent B1500A User’s Guide, Edition 13 3- 17
Installation
Installing Accessories
To Connect ASU
Agilent E5288A Atto Sense and Switch Unit (ASU) adds the following functions to
the B1500A. To use the ASU, it must be connected to the SMU which is connected
together when the calibration is performed.
Atto ampere (aA) level current measurement capability
Automatic switching of ASU input resource connected to ASU output
NOTE For the installation of the ASU and the connection to the DUT interface (probe card,
manipulators, and so on), contact your favorite prober vender. The prober vender
will have the solutions. Dimensions of the ASU are 132 mm (W)
88.5 mm (H)
50 mm (D) excluding the connectors.
Figure 3-1 Image of ASU Installation
Required parts Agilent 16495K Connector plate
Agilent 16493M ASU cable
Agilent N1300A CMU cable or Agilent 16048D/E test leads
Agilent N1254A-108 Magnet stand
BNC cable, coaxial cable, manipulator, wire, and so on
To install 16495K Fix the 16495K plate to the shielding box or something that will cover the DUT
interface. See Agilent 16495 Installation Guide. The 16495K is the plate that has the
mechanism to block the light from the cable hole used to pass the cables in the
shielding box.
6KLHOGLQJER[
7R%$&08
. .
$68
'87LQWHUIDFH
$68
3UREHU
&085HWXUQ
&085HWXUQ