3- 16 Agilent B1500A User’s Guide, Edition 13
Installation
Installing Accessories
Table 3-4 To Connect Connector Plate
Cables
Connector
Plate
B1500A Terminals
Agilent 16493J Interlock Cable, 3 m or 1.5 m Intlk Interlock
Agilent 16493L GNDU Cable, 3 m or 1.5 m GNDU GNDU
Agilent 16493K
a
Kelvin Triaxial Cable, 3 m or
1.5 m, for the Kelvin connection
16495F provides the inputs 1 to 12 (6 Kelvin).
16495H provides the inputs 1 to 6 (3 Kelvin).
16495J provides the inputs 1 to 8 (4 Kelvin).
a. Use Agilent 16494B Kelvin Triaxial Cable to connect to Agilent E5250A
Low Leakage Switch Mainframe or the connector plate for Agilent 4142B
Modular DC Source/Monitor by the Kelvin connection.
odd number
SMU
Force
even number Sense
Agilent 16494A Triaxial Cable, 3 m or 1.5 m,
for non-Kelvin connection
16495H provides the inputs 1 to 6.
16495J provides the inputs 1 to 8.
odd number
or
even
number
SMU Force
Agilent 16493P SPGU Cable, 3 m or 1.5 m
16495H provides the inputs 7 to 12.
16495J provides the inputs 9 to 12.
odd number
or
even
number
SPGU Output 1
Output 2
Agilent N1300A CMU Cable
b
, 3 m or 1.5 m
16495H provides the inputs 7 to 12.
16495J provides the inputs 9 to 12.
b. To extend the four-terminal pair (4TP) connection to the connector plate
inputs, connect the Hcur and Hpot cables together and the Lcur and Lpot
cables together by using the BNC T-type adapter (female-female-male, Agi
-
lent part number 1250-2405), then connect them to the connector plate inputs.
To extend the 4TP connection to the DUT interface, connect the cables to the
connector plates input connectors individually, and connect the Hcur and
Hpot terminals together and the Lcur and Lpot terminals together on the DUT
interface. See Figure 3-21.
It is ok to leave open the ground wire extended from the CMU cable.
odd number
or
even
number
CMU
Hcur/Hpot
Lcur/Lpot