Agilent Technologies B1500A Guide de l'utilisateur Page 70

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 202
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 69
2- 28 Agilent B1500A User’s Guide, Edition 13
Introduction
Measurement Modules
HRSMU - High Resolution SMU
This section describes typical specification of the high resolution source/monitor
unit (HRSMU). There is no restriction for the number of modules installed in one
mainframe.
Maximum voltage, current, output power: ± 100 V, ± 100 mA, 2 W
Minimum range: 0.5 V, 10 pA (1 pA when the E5288A ASU is used)
Output/measurement value and resolution: see Table 2-11 to Table 2-14.
Figure 2-6 HRSMU Output and Measurement Ranges











9ROWDJH9
&XUUHQWP$

Vue de la page 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 201 202

Commentaires sur ces manuels

Pas de commentaire