Agilent B1500A User’s Guide, Edition 13 4- 3
Specifications
Hardware
Hardware
Specification Conditions
The measurement and output accuracy are specified at the rear panel connector
terminals when referenced to the Zero Check terminal. The B1530A WGFMU
measurement and output accuracy are specified at the output terminal of the RSU.
Accuracy is specified under the following conditions:
1. Temperature: 23 C ± 5 C
2. Humidity: 20 % to 60 %
3. After 40 minutes warm-up followed by self-calibration
4. Ambient temperature change less than ± 1 C after self-calibration execution,
not applicable for MFCMU and WGFMU
5. Measurement made within one hour after self-calibration execution,
not
applicable for MFCMU and WGFMU
6. Calibration period: 1 year
7. SMU integration time setting:
1 PLC (1 nA to 1 A range, voltage range)
20 PLC (100 pA range)
50 PLC (1 pA to 10 pA range)
Averaging of high speed ADC: 128 samples per 1 PLC
8. SMU filter: ON for HPSMU, MPSMU, and HRSMU
9. SMU measurement terminal connection: Kelvin connection
10. WGFMU load capacitance: 25 pF or less
NOTE Agilent Technologies is responsible for removing, installing, and replacing the
B1500A modules. Contact your nearest Agilent Technologies to install and calibrate
the B1500A modules.