Agilent B1500A User’s Guide, Edition 13 3- 45
Installation
Connecting Measurement Devices
Connections for High Current Measurements
When you force or measure a large current, you may want to use a Kelvin
connection (4-wire, remote sensing) to eliminate the residual resistance effects of
test leads and contacts. For example, you can use the following connections as
Kelvin connections on the test fixture.
Examples: Kelvin Connection
To cancel the effects of the residual resistance, test leads must be connected as close
as possible to the DUT.
6089 $0D[
6HQVH)RUFH
*XDUG *XDUG
6HQVH)RUFH
*XDUG *XDUG
6HQVH)RUFH
*XDUG *XDUG
6HOHFWHG
*DWH
6RXUFH
'UDLQ
6XEVWUDWH