4- 24 Agilent B1500A User’s Guide, Edition 13
Specifications
Hardware
Measurement accuracy
Measurement accuracy at the measurement port of the MFCMU or the MFCMU
cable (1.5 m or 3.0 m)
|Z| accuracy: ± E (%)
accuracy: ± E/100 (radian)
C accuracy:
± E (%) at Dx 0.1
± E(1+Dx
2
)
1/2
(%) at Dx > 0.1
D accuracy:
± E/100 at Dx 0.1
± E(1+Dx)/100 at Dx > 0.1
G accuracy:
± E/Dx (%) at Dx 0.1
± E(1+Dx
2
)
1/2
/Dx (%) at Dx > 0.1
where,
Zx: Impedance measurement value ()
Dx: Measurement value of D
E = E
P
’ + (Z
S
’ / |Zx| + Y
O
|Zx| ) 100 (%)
E
P
’ = E
PL
+ E
POSC
+ E
P
(%)
Y
O
’ = Y
OL
+ Y
OSC
+ Y
O
(S)
Z
S
’ = Z
SL
+ Z
OSC
+ Z
S
()
For the parameter value, see Table 4-11 to Table 4-13. Calculation example of
the measurement accuracy is also shown in Table 4-14.
Specification conditions:
Temperature: 23 C ± 5 C
Integration time: 1 PLC or 16 PLC