Triggering 4
InfiniiVision Oscilloscopes User’s Guide 133
MIL-STD 1553 Triggering
To set up the oscilloscope to capture MIL-STD 1553 signals, see “Setup for MIL-STD
1553 Signals” on page 132.
To set up a MIL-STD 1553 trigger:
1 In the front panel Trigger section, press [More].
2 In the More Trigger Menu, press Trigger; then, select the MIL-1553 trigger type.
3 Press Settings.
4 In the MIL-STD-1553 Trigger Settings Menu, press Trigger; then, select from these
options:
Data Word Start – triggers on the start of a Data word (at the end of a valid Data
Sync pulse).
Data Word Stop – triggers on the end of a Data word.
Command/Status Word Starttriggers on the start of Command/Status word (at
the end of a valid C/S Sync pulse).
Command/Status Word Stop – triggers on the end of a Command/Status word.
Remote Terminal Address – triggers if the RTA of the Command/Status word
matches the specified value.
When this option is selected, the RTA softkey appears and lets you select the hex
Remote Terminal Address value to trigger on. If you select 0xXX (don’t cares), the
oscilloscope will trigger on any RTA.
Remote Terminal Address + 11 Bits – triggers if the RTA and the remaining 11 bits
match the specified criteria.
When this option is selected, the RTA + Bits softkey appears and opens the
MIL-STD-1553 RTA + 11 Bits Menu. In this menu:
The RTA softkey lets you select the hex Remote Terminal Address value.